Abstract
Objective: We investigated the different in acute hemodynamic changing between the Quick-Opt Method Programming versus device settings and Echocardiographic Optimization Programming. Method: After 01 month, 19 patients was implanted the St. Jude devices with Quick-Opt Programming who was assessed by echocardiographic the hemodynamic parameters of programming with the device settings, Quick-Opt and echocardiographic optimization. Results: We found no significant different of programming between the device settings versus Quick-Opt in AV delay (170 ms vs 181,67± 9,37 ms; p=0,45); PV delay (150 ms vs 133,33±14,35; p=0,08), but there is significant different in VV delay (20 ms vs 43,12±14,16 ms; p=0,006). Echocardiography didn't show the different of acute hemodynamic changing between the Quick Opt method programming versus device settings (p>0,05). We found no significant different of programming between the device settings versus Quick-Opt in AV delay (177,50±16,02 vs 181,67± 9,37 ms; p=0,51); PV delay (142,50±12,15 ms vs 133,33±14,35; p=0,09); VV delay (30,42±10,75 ms vs 43,12±14,16 ms; p=0,06). The echocardiographic optimization programming is better than the Quick Opt method programming in acute hemodynamic changing (p<0,05). Conclusion: The Quick Opt method programming is the equal of device settings but worse than echocardiographic optimization programming in acute hemodynamic changing.